タイトル |
|
|
タイトル |
GAIT CYCLE DETECTION USING A TRI-AXIAL ACCELEROMETER AND A GYROSCOPE IN HEMIPLEGIC PATIENTS : A PRELIMINARY REPORT |
|
言語 |
en |
言語 |
|
|
言語 |
eng |
主題 |
|
|
言語 |
en |
|
主題Scheme |
Other |
|
主題 |
functional electrical stimulation (FES) |
主題 |
|
|
言語 |
en |
|
主題Scheme |
Other |
|
主題 |
gait cycle detection |
主題 |
|
|
言語 |
en |
|
主題Scheme |
Other |
|
主題 |
hemiplegia |
主題 |
|
|
言語 |
en |
|
主題Scheme |
Other |
|
主題 |
accelerometer |
資源タイプ |
|
|
資源タイプ識別子 |
http://purl.org/coar/resource_type/c_6501 |
|
資源タイプ |
journal article |
アクセス権 |
|
|
アクセス権 |
open access |
|
アクセス権URI |
http://purl.org/coar/access_right/c_abf2 |
作成者 |
Tomite, Takenori
Shimada, Yoichi
Matsunaga, Toshiki
Sasaki, Kana
Yoshikawa, Takayuki
Iwami, Takehiro
|
内容記述 |
|
|
内容記述タイプ |
Abstract |
|
内容記述 |
Abstract Objective : To evaluate the capability for gait cycle detection using a tri-axial accelerometer and gyroscope in hemiplegic patients. Materials & Methods : Twenty hemiplegic patients participated in this study. The sensors were placed on the tibial tubercle of the affected knee. We divided the patients into groups according to Brunnstrom stage to evaluate whether the sensors can detect gait cycle irrespective of the degree of paralysis. To evaluate whether errors and delay times seen in signals of the sensors were too pronounced for a hemiplegic patient’s gait, we asked 5 hemiplegic patients, who had errors and delay times, to walk with functional electrical stimulation (FES) from signals of only the sensors and compared walking speed and step cadences for walking with and without stimulation. Result : Outputs of the sensor signals had some errors and were behind the output of heel sensor signals. The total number of steps was 912. The total number of errors was 20 (2.0%). Average delay time was 0.058 sec (N=20). There were no significant differences among Brunnstrom stages in terms of appearances of errors and average delay times (p>0.05 ; Kruskal-Wallis rank-sum test). Five patients who were asked to walk with FES from signals of the sensors had faster walking speed and fewer steps than when walking without FES (p<0.05 ; Wilcoxon signed rank test). Conclusion : Although errors and delay times were observed in the output of the sensor signals, patients who were asked to walk with FES from the sensor signals could obtain a better walking ability. |
|
言語 |
en |
出版タイプ |
|
|
出版タイプ |
VoR |
|
出版タイプResource |
http://purl.org/coar/version/c_970fb48d4fbd8a85 |
書誌情報 |
ja : 秋田医学
巻 38,
号 3/4,
p. 105-110,
発行日 2012-03-01
|
収録物識別子 |
|
|
収録物識別子タイプ |
ISSN |
|
収録物識別子 |
03866106 |
収録物識別子 |
|
|
収録物識別子タイプ |
NCID |
|
収録物識別子 |
AN00009294 |
出版者 |
|
|
出版者 |
秋田医学会 |
|
言語 |
ja |