{"created":"2023-07-25T10:25:31.507513+00:00","id":5905,"links":{},"metadata":{"_buckets":{"deposit":"b4010da9-b154-48e1-a2a5-6bd384ce67bd"},"_deposit":{"created_by":3,"id":"5905","owners":[3],"pid":{"revision_id":0,"type":"depid","value":"5905"},"status":"published"},"_oai":{"id":"oai:air.repo.nii.ac.jp:00005905","sets":["611:939:940"]},"author_link":["17275","17277","17278","17279","17274","17280","17276"],"item_10001_biblio_info_7":{"attribute_name":"書誌情報","attribute_value_mlt":[{"bibliographicIssueDates":{"bibliographicIssueDate":"2021","bibliographicIssueDateType":"Issued"},"bibliographicIssueNumber":"4","bibliographicVolumeNumber":"21","bibliographic_titles":[{"bibliographic_title":"SENSORS"}]}]},"item_10001_description_5":{"attribute_name":"内容記述(抄録)","attribute_value_mlt":[{"subitem_description":"This study aimed to evaluate the capability of a piezoelectric sensor to detect a heart murmur in patients with congenital heart defects. Heart sounds and murmurs were recorded using a piezoelectric sensor and an electronic stethoscope in healthy neonates (n = 9) and in neonates with systolic murmurs caused by congenital heart defects (n = 9) who were born at a hospital. Signal data were digitally filtered by high-pass filtering, and the envelope of the processed signals was calculated. The amplitudes of systolic murmurs were evaluated using the signal-to-noise ratio and compared between healthy neonates and those with congenital heart defects. In addition, the correlation between the amplitudes of systolic murmurs recorded by the piezoelectric sensor and electronic stethoscope was determined. The amplitudes of systolic murmurs detected by the piezoelectric sensor were significantly higher in neonates with congenital heart defects than in healthy neonates (p < 0.01). Systolic murmurs recorded by the piezoelectric sensor had a strong correlation with those recorded by the electronic stethoscope (rho = 0.899 and p < 0.01, respectively). The piezoelectric sensor can detect heart murmurs objectively. Mechanical improvement and automatic analysis algorithms are expected to improve recording in the future.","subitem_description_type":"Other"}]},"item_10001_publisher_8":{"attribute_name":"出版者","attribute_value_mlt":[{"subitem_publisher":"MDPI"}]},"item_10001_relation_14":{"attribute_name":"DOI","attribute_value_mlt":[{"subitem_relation_type":"isIdenticalTo","subitem_relation_type_id":{"subitem_relation_type_id_text":"10.3390/s21041376","subitem_relation_type_select":"DOI"}}]},"item_10001_relation_25":{"attribute_name":"関連リンク","attribute_value_mlt":[{"subitem_relation_name":[{"subitem_relation_name_text":"http://dx.doi.org/10.3390/s21041376"}],"subitem_relation_type_id":{"subitem_relation_type_id_text":"http://dx.doi.org/10.3390/s21041376","subitem_relation_type_select":"DOI"}}]},"item_10001_rights_15":{"attribute_name":"著作権等","attribute_value_mlt":[{"subitem_rights":"© 2021 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access\narticle distributed under the terms and conditions of the Creative Commons Attribution\n(CC BY) license (http://creativecommons.org/licenses/by/4.0/)."}]},"item_10001_source_id_9":{"attribute_name":"ISSN","attribute_value_mlt":[{"subitem_source_identifier":"1424-8220","subitem_source_identifier_type":"ISSN"}]},"item_10001_version_type_20":{"attribute_name":"著者版フラグ","attribute_value_mlt":[{"subitem_version_resource":"http://purl.org/coar/version/c_970fb48d4fbd8a85","subitem_version_type":"VoR"}]},"item_creator":{"attribute_name":"著者","attribute_type":"creator","attribute_value_mlt":[{"creatorNames":[{"creatorName":"Takahashi, Kiichi"}],"nameIdentifiers":[{}]},{"creatorNames":[{"creatorName":"Ono, Kyoichi"}],"nameIdentifiers":[{}]},{"creatorNames":[{"creatorName":"Arai, Hirokazu"}],"nameIdentifiers":[{}]},{"creatorNames":[{"creatorName":"Adachi, Hiroyuki"}],"nameIdentifiers":[{}]},{"creatorNames":[{"creatorName":"Ito, Masato"}],"nameIdentifiers":[{}]},{"creatorNames":[{"creatorName":"Kato, Akie"}],"nameIdentifiers":[{}]},{"creatorNames":[{"creatorName":"Takahashi, Tsutomu"}],"nameIdentifiers":[{}]}]},"item_files":{"attribute_name":"ファイル情報","attribute_type":"file","attribute_value_mlt":[{"accessrole":"open_date","date":[{"dateType":"Available","dateValue":"2023-02-23"}],"displaytype":"detail","filename":"iA_2022_74.pdf","filesize":[{"value":"16.8 MB"}],"format":"application/pdf","licensetype":"license_note","mimetype":"application/pdf","url":{"label":"iA_2022_74.pdf","url":"https://air.repo.nii.ac.jp/record/5905/files/iA_2022_74.pdf"},"version_id":"5f56603d-6eca-4b32-bd95-d45fc1a3936e"}]},"item_keyword":{"attribute_name":"キーワード","attribute_value_mlt":[{"subitem_subject":"cardiac murmur","subitem_subject_scheme":"Other"},{"subitem_subject":" congenital heart defects","subitem_subject_scheme":"Other"},{"subitem_subject":" heart sound","subitem_subject_scheme":"Other"},{"subitem_subject":" noninvasive","subitem_subject_scheme":"Other"},{"subitem_subject":" piezoelectric sensor","subitem_subject_scheme":"Other"}]},"item_language":{"attribute_name":"言語","attribute_value_mlt":[{"subitem_language":"eng"}]},"item_resource_type":{"attribute_name":"資源タイプ","attribute_value_mlt":[{"resourcetype":"journal article","resourceuri":"http://purl.org/coar/resource_type/c_6501"}]},"item_title":"Detection of Pathologic Heart Murmurs Using a Piezoelectric Sensor","item_titles":{"attribute_name":"タイトル","attribute_value_mlt":[{"subitem_title":"Detection of Pathologic Heart Murmurs Using a Piezoelectric Sensor"}]},"item_type_id":"10001","owner":"3","path":["940"],"pubdate":{"attribute_name":"公開日","attribute_value":"2023-02-23"},"publish_date":"2023-02-23","publish_status":"0","recid":"5905","relation_version_is_last":true,"title":["Detection of Pathologic Heart Murmurs Using a Piezoelectric Sensor"],"weko_creator_id":"3","weko_shared_id":-1},"updated":"2023-07-25T10:39:33.912443+00:00"}